Highlights

The N7764C four-channel variable optical attenuators for single-mode fiber enables highly efficient optical transceiver and network integration testing. It is based on electrically controlled variable optical attenuator (VOA) modules.

Applications

  • Optical transceiver testing
  • Optical network integration testing

Power Setting and Control Modes

In power setting mode, which is available on all N77-C variable optical attenuators, the power level at the attenuator output is set. The N7764C uses the feedback signal from a photodiode after a monitor tap to control the desired power level at the optical output.

When power control is active, the instrument also automatically corrects for power changes at the input so that the output power level you set is maintained with ±0.02 dB typical repeatability.

Attenuation setting mode

In attenuation mode, the calibrated value of attenuation in dB is set directly, without use of the power monitor. This supports use with legacy software written for attenuators without power monitors or when the power monitor reading is impaired, as due to modulation or from a signal at another wavelength or in the reverse direction.

Advanced control of power change

Especially for testing devices that are sensitive to rapid power changes, the rate of change can be set on the instrument and applies to the attenuation mode. The N7764C can be set from 0.1 to 1000 dB/s. New with the N77-C attenuators is the ability to program a ramp of attenuation steps, including dwell time at each step, which can be either stepped automatically or synchronized with an external trigger. Trigger outputs are also available. This new functionality can be used instead of sending individual program commands for each step and aid in efficient synchronization with other instruments. In power control mode, during attenuation sweeps, as well as when activating the shutter mode, 1000 dB/s change rate is applied.

Calibration offsets for external losses

Comprehensive offset functionality allows you to calibrate the optical path in various test set-ups. There is an offset for the attenuation factor, and an independent offset for the output power level, to calibrate for losses due to patch cords, connectors, and switches. Additionally, wavelength and offset value pairs can be stored in a table to compensate for wavelength-dependent effects in the optical path of the set-up. This allows you to precisely set the optical power level directly at the input interface of your device under test.

New user attenuation-mode calibration

To calibrate the attenuation setting mode for a chosen source wavelength, a new self-calibration has been implemented. The calibration process, executed at the chosen wavelength and power higher than 0 dBm, and with the setup kept stable, will run for about 2 minutes. The calibration data can be reenabled after a preset or reboot, or the configuration can be saved with enabled calibration.

Key Specifications

Attenuation Range
0 dB to 40 dB
Fiber Mode
Single-mode
Insertion Loss Choice
<2.2 dB at 1550 nm
Mainframe Compatibility
Standalone
Number of Channels
4
Power Meter Ports
n/a
Attenuation Range
Fiber Mode
Insertion Loss Choice
Mainframe Compatibility
Number of Channels
Power Meter Ports
0 dB to 40 dB
Single-mode
<2.2 dB at 1550 nm
Standalone
4
n/a
View More
Attenuation Range:
0 dB to 40 dB
Fiber Mode:
Single-mode
Insertion Loss Choice:
<2.2 dB at 1550 nm
Mainframe Compatibility:
Standalone
Number of Channels:
4
Power Meter Ports:
n/a
Power Setting Mode:
-50 dBm to +20 dBm
Type:
Optical Attenuator
Wavelength Range:
1260 nm to 1640 nm
N7764C 4-Channel Optical Attenuator