L4411A Firmware Revision History
Changes from Revision 2.35 to 2.39
Defects resolved in this release:
- Rearranged the ordering of the learn string (*LRN?) to restore the instrument state after queried by *LRN?
- Temperature setting changes will not default the MATH states or reading memory unless Temperature (and the Transducer type) are active
- Intermittent Freq/Period measurement timeouts resolved
- Added measurement capability for RTDs other than 100 ohms
- Transducer settings correctly restored following *RCL
- Eliminated unnecessary settling delays to increase Offset Compensation (OCOMP) speed
- Corrected calibration data and state storage issue involving instrument non-volatile memory
Changes from Revision 2.24 to 2.35
Defects resolved in this release:
- Minor changes made for production test. No change to the user.